Troubleshooting guide

Troubleshooting steps from the EAGLE, LION, ANT, HAWK and BetonicTCP manuals. Search for the symptom or the on-screen message; if it is not solved, open a service request.

10 entries

The indicator does not turn onEAGLELION
Windows shows a startup error (KRPC models with built-in computer)EAGLELION

Possible cause

Cutting the power suddenly can cause data loss and hardware failures.

Steps to solve it

  1. If Windows shows an error while starting, use Startup Repair.

Note KRPC models contain a computer, so the power must not be cut suddenly when shutting down. Use Start > Shut down in Windows, or press the ON button on the rear panel of the indicator once, briefly.

The weight on the indicator display keeps going negativeEAGLELION

Possible cause

One or more load cells of the scale are faulty.

Steps to solve it

  1. Contact technical support to locate the faulty load cell.

Note The health test of the connected load cell can be done in the IC SAYAÇ (internal counter) menu of the indicator. On LION this screen should read a value between 0 and 167772.

The indicator display goes darkEAGLELION

Possible cause

The scale capacity is lower than the weight placed on it.

Steps to solve it

  1. Change the capacity of that scale in the Betosistem software.
  2. Since the capacity has changed, repeat the calibration.
Stock shortfalls keep appearingEAGLELIONBetonicTCP

Steps to solve it

  1. Check the scale calibrations at least once a month.
  2. Before calibrating, clean around the scale and the canvas under the discharge gates.
  3. Make sure the scale hangs level by keeping the metal rods (studs) that connect the load cells to the scale at equal lengths.
  4. If the shortfall continues, place a 50 kg weight on each corner where a load cell is attached, one at a time. The indicator must show the same value at every corner.
  5. If any corner differs by even (+/-) 1 kg, check the load cell at that corner.
  6. If there is no calibration error, check the manual productions and material intakes in BetonicTCP under Program settings > Manuals.
  7. If the shortfall still continues, call service for technical support.
The software and the indicator do not communicate (the serial port indicator in the bottom right corner is not spinning)EAGLELIONBetonicTCP

Possible cause

If the serial port status indicator in the bottom right corner of the software is not spinning, there is a communication problem.

Steps to solve it

  1. Check the serial port settings in the software.
  2. In Device Manager > Mice and other pointing devices, if there is a device named "Microsoft Serial Ballpoint", turn it off with "Disable".
  3. If the problem persists, call an authorised service.
Indicator weights do not appear in the software (after first installation)EAGLELIONBetonicTCP

Possible cause

The plant parameters have not been sent.

Steps to solve it

  1. Send all parameters in order from Files > M3 > Plant settings.
The scale gates always appear open in the softwareEAGLELION

Steps to solve it

  1. Make sure cable no. 900 is plugged into the Serial 2 connector.
  2. Check the power supply of the PLC.
  3. If the gates appear open and you also cannot see the scale weights on the screen, call service.
The mimic panel sounds an intermittent siren and the LEDs flashEAGLELION

Steps to solve it

  1. Make sure cable no. 900 is plugged into the Serial 2 connector.
  2. Check the power supply of the PLC.
  3. If it continues, call service.
Calibration settings return to the old values after a power cutEAGLE

Possible cause

The values were skipped on the "SUrE" (confirm) screen without confirming. New values are then kept only until the power is cut; after a power cut the indicator returns to the previous settings.

Steps to solve it

  1. Repeat the calibration and confirm on the "SUrE" (confirm) screen at the end. Confirmed values are kept even after a power cut.

This information is taken from our user manuals. Call the technical support line before carrying out any step you are unsure about.