Weighing is wrong after C 01, C 02 or C 03 was changedANTHAWK

Possible cause

Scale capacity and calibration are set by C 01 (division value), C 02 (number of divisions) and C 03 (decimals); changing them breaks the calibration.

Steps to solve it

  1. Restore the old values or recalibrate for the new values.
  2. Calibrate with reference (test) weights of known mass.

Note In case of a fault, the indicator or load cells can be tested in the IC SAYAÇ (internal counter) menu.

This information is taken from our user manuals. Call the technical support line before carrying out any step you are unsure about.